Cuvettes are essential components in spectroscopic analysis. Selecting the wrong cuvette can lead to inaccurate readings, poor reproducibility or even total measurement failure.
This guide explains the difference between quartz and glass cuvettes, helping you choose the right one for your UV/Vis spectrophotometer.

What Is a Cuvette and Why Does Material Matter?

A cuvette is a small, transparent cell used to hold liquid samples for absorbance or fluorescence measurements.

Its optical properties determine the accuracy of your results, including transmission range, refractive index and chemical resistance.

Different materials transmit different wavelength ranges. Choosing the right cuvette ensures that your spectrophotometer can properly detect the light passing through your sample.

 

Wavelength Transmission Comparison

Property Quartz Cuvette Optical Glass Cuvette
Usable Range 190 – 900 nm 360 – 900 nm
UV Transparency Excellent Limited
Hardness & Durability Very high (scratch-resistant) Moderate
Chemical Resistance Excellent (acid/base resistant) Good
Cost Higher Lower

 

Key point: If your experiment involves UV wavelengths below 360 nm, only quartz cuvettes can provide accurate results.


Why Quartz Outperforms Glass?

Quartz cuvettes have a broader optical window from 120 nm to 4500 nm, allowing precise measurement across UV, visible, and NIR regions.

In contrast, glass cuvettes only transmit 400 – 4000 nm and exhibit multiple ionic absorption peaks. This means quartz delivers more accurate, stable, and reproducible data for scientific and industrial applications.

How to Identify Quartz vs. Glass Cuvettes?

Even though quartz and glass look similar, they differ significantly in hardness and refractive index.

Method 1: Label Identification

Quartz cuvettes are typically marked with a “Q” and glass with a “G.”

If no mark exists, test transmission in the UV region: quartz shows higher transmittance below 350 nm.

Method 2: Spectrophotometer Test

  1. Set your spectrophotometer to 250 nm.

  2. Zero the instrument.

  3. Insert an empty cuvette:

  • Absorbance < 0.07 Abs → Quartz
  • Absorbance > 0.07 Abs → Glass or unclean quartz

 

Tip: Always clean cuvettes thoroughly! Residues can mimic glass like absorbance.

 

When to Use Each Material

Application Recommended Cuvette
UV/Vis spectroscopy (190 – 900 nm) Quartz
Visible-only measurements (> 400 nm) Glass
Fluorescence or photoluminescence Quartz
Cost-sensitive educational labs Glass
High-precision analytical chemistry Quartz

 

Other Factors to Consider

  • Path length: Commonly 10 mm; shorter/longer options affect sensitivity.
  • Shape: Standard square vs. round (for flow or fluorescence).
  • Chemical compatibility: Quartz withstands acids, bases and solvents better than glass.
  • Maintenance: Avoid touching optical windows; rinse with solvent and store dust-free.

 

Looking for high-quality quartz cuvettes for your UV/Vis analysis?

Explore J&K Scientific Cuvettes ›

Need technical advice on selecting cuvettes?

Contact Our Technical Team ›

 